Specifications
|
Instrument Configuration
|
|
Outer Dimensions:
|
length
|
x: 160 cm (63”)
|
x: 160 cm (63”)
|
|
|
depth
|
y: 70 cm (28”)
|
y: 70 cm (28”)
|
|
|
height
|
z: 72 cm (28”)
|
y: 70 cm (28”)
|
|
|
weight
|
w: 100-126 kg (220-278 lbs)
|
w: 100-130 kg (200-287 lbs)
|
|
|
|
|
|
|
Working Dimensions:
|
length
|
x: 130 cm (51”)
|
x : 125 cm (49”)
|
|
|
depth
|
y: 37 cm (15”)
|
y: 35 cm (14”)
|
|
|
height
|
z: 11.5 cm (4.5”)
|
z: 13.6 cm (5.4”)
|
|
|
|
|
|
|
Deck Capacity
|
|
36 microplate positions
|
36 microplate positions
|
|
Arms
|
|
1 or 2 arms
|
1 or 2 arms
|
|
Channels per Arm
|
|
2, 4 or 96 channels
|
4, 8 or 96 channels
|
|
Instrument Power
|
|
400 VA
|
400 VA
|
|
|
|
|
|
|
Platform Specs - All Types
|
|
|
|
|
Software, Operating System
|
Lirix Software Package, Windows™ XP, Service Pack 2 or higher
|
|
|
|
|
|
|
|
|
|
|
Computer
|
- Embedded in pipetting deck, data access via USB 2.0 port in support stand - Technical data: CeleronM 1.5GHz, 512 MB RAM, >40 GB HD - Communication: 2 serial ports RS-232, 4 USB 2.0 ports, 1 parallel port, LAN (RJ-45)
|
|
Communication
|
High speed CAN bus; max speed of 1MBit/sec; 2-6 CAN ports in support stands
|
|
|
|
|
|
|
|
|
|
|
Position Accuracy
|
in X/Y/Z: +/- 0.2 mm
|
|
Plate Handler “Plate-GripX”
|
Lifts objects up to 400 g (0.9 lbs)
|
|
|
|
|
|
|
|
|
|
|
Liquid Detection with Pressure Sensor
|
- Single or multiple detection, independent for each channel - Liquid level detection and clot detection
|
|
Tip Types
|
Disposable tips with and without filters; channels with independent Y and Z movement
|
|
|
|
|
Pump Sizes
|
Medium pumps:
|
250 µl
|
Large pumps:
|
1000 µl
|
|
Volume Ranges
|
Medium tips:
|
2-250 µl
|
Large tips:
|
10-1000 µl
|
|
|
|
|
|
|
|
Precision and Accuracy
|
Pump Size
|
Volume dispensed
|
Precision (CV%)
|
Inaccuracy (%)
|
|
|
250 µl
|
20 µl
|
1.0 %
|
2.5 %
|
|
|
|
200 µl
|
0.5 %
|
1.0 %
|
|
|
1’000 µl
|
100 µl
|
1.0 %
|
3.0 %
|
|
|
|
1’000 µl
|
0.5 %
|
1.0 %
|
|
|
|
|
|
|
|
|
|
|
Container Format
|
Lirix supports all liquid container formats up to 384-well plates
|
|
Pipetting Skills
|
- Single pipetting, multi pipetting, parallel pipetting, aliquoting, archiving - Serial dilutions, pre-dilutions, multiple common pre-dilutions - Liquid mixing in source and destination container
|
|
|
|
|
|
|
|
|
|
|
Plate Handler “Plate-GripX”
|
Gripper fingers are manipulated with two tip adapters
|
|
Barcode Scanner “Tube-ScanX”
|
Most common barcode types (e.g. Code 39, Interleaved 2 of 5, Codabar, Code 128, etc.)
|
|
Plate Shaker “Plate-ShakeX”
|
Linear and orbital shaking mode with speed range 100-2’000 rpm and an amplitude of 2 mm
|
|
Heating / Cooling Devices
|
Temperature control devices for microplates, tubes and customized racks and containers.
|
|
SPE Vacuum Station
|
Vacuum-based solid phase extraction device for NA and protein isolation.
|
|
Magnetic Bead Station
|
Magnetic bead separation for NA and protein isolation.
|
|
|
|
|
|
|
|
|
|
|
Voltage, Frequency, Power
|
Embedded in pipetting deck; 110/230 VAC; 50/60 Hz
|
|
|
|
|
|
|
|
|
|
|
Standards
|
In compliance with CE standards (bears the CE mark)
|
|
|
|
|
|
|
|
|
|
|
Temperature
|
Operating: 15-32°C (59-90°F); Storage: -10-60°C (14-140°F)
|
|
Relative Humidity
|
Operating and storage: 30-80 % rH (non-condensing) at 30°C (86°F)
|
|
|
|
(C) 2007 - All rights reserved |
|